Table 4. Features and Controls (cont.) 
Item Description 
 
AC Adapter/Charger Input Terminal 
 
Retractable Lens Cover  
 
Torch/Flashlight (not applicable for Ti100) 
 
Visual Camera (not applicable for Ti100) and Lens 
 
Infrared Camera Lens 
 Laser 
Pointer 
 Secondary 
Trigger 
 Primary 
Trigger 
 
Hand Strap Anchor Post (Right and Left) 
 
IR-OptiFlex™ Focus Control (not applicable for Ti100, Ti105, 
TiR105) 
 Lithium-ion 
Smart Battery 
 
2-Bay Battery Charging Base 
 
AC Power Adapter with Mains Adapters 
Focus 
Models Ti110, Ti125, TiR110, and TiR125 have IR-OptiFlex focus. The 
Ti100, Ti105, and TiR105 use a large depth of field focus-free system. Models 
with IR-OptiFlex focus can operate in a focus-free mode, but also have the 
flexibility for close focus situations (<122 cm/48 inches) with a one-touch, 
fine-tune focus capability. 
The focus-free system can focus at a minimum distance of 122 cm (48 in) and 
beyond with no adjustment.  
Correct focus is important in all imaging applications. Correct focus makes 
sure that the infrared energy is correctly directed onto the pixels of the 
detector. Without the correct focus, the thermal image can be blurry and the 
radiometric data will be inaccurate. Out-of-focus infrared images are 
frequently unusable or of little value.