Fluke Ti100 , 9 Hz thermography camera, -20 to +250 °C, 160 x 120 pix bolometer matrix 3810290 User Manual

Product codes
3810290
Page of 56
Ti100, Ti105, Ti110, Ti125, TiR105, TiR110, TiR125 
Users Manual 
18 
Temperature Measurement 
All objects radiate infrared energy. The quantity of energy radiated is based on 
the actual surface temperature and the surface emissivity of the object. The 
Imager senses the infrared energy from the surface of the object and uses this 
data to calculate an estimated temperature value. Many common objects and 
materials such as painted metal, wood, water, skin, and cloth are very good at 
radiating energy and it is easy to get relatively accurate measurements. For 
surfaces that are good at radiating energy (high emissivity), the emissivity 
factor is 
90 % (or 0.90). This simplification does not work on shiny surfaces 
or unpainted metals as they have an emissivity of 
<
0.60. These materials are 
not good at radiating energy and are classified as low emissivity. To more 
accurately measure materials with a low emissivity, an emissivity correction is 
necessary. Adjustment to the emissivity setting will usually allow the Imager 
to calculate a more accurate estimate of the actual temperature. 
 Warning 
To prevent personal injury, see emissivity information for 
actual temperatures. Reflective objects result in lower than 
actual temperature measurements. These objects pose a 
burn hazard. 
Note 
Surfaces with an emissivity 
<
0.60 make reliable and consistent 
determination of actual temperatures problematic. The lower the 
emissivity, the more potential error is associated with the Imager's 
temperature measurement calculations, even when emissivity and 
reflected background adjustments are attempted and performed 
properly. 
More information is available on emissivity. We recommend the study of this 
topic to get the most accurate temperature measurements. 
SmartView
 Software 
SmartView
®
 software is supplied with the Imager. This software is intended 
for Fluke Imagers and contains features to analyze images, organize data and 
information, and make professional reports. SmartView
®
 allows audio 
annotations and IR-PhotoNotes to be reviewed on a PC. SmartView
®
 is used to 
export IR and visible images as .jpeg, .jpg, .jpe, .jfif, .bmp, .gif, .dip, .png, .tif, 
or .tiff formatted files.