5 First steps 
18 
2.
  Turn the mounting ring anticlockwise and remove the protection 
glass. 
After mounting/removing the protection glass, activate or deactivate 
the 
Protection glass
 option, see Lens, page 10. If this option is set 
incorrectly, the specified measuring accuracy is not ensured. 
 
Mounting/removing high-temperature filter 
Installation: 
1.
  Attach the high-temperature filter (with red fitting) fastened to 
the red mounting ring to the lens and turn the mounting ring 
clockwise to the stop. 
2.
  Remove the red mounting ring from the high-temperature filter. 
Removal: 
1.
  Attach the red mounting ring to the high-temperature filter. 
2.
  Turn the mounting ring anticlockwise and remove the high-
temperature filter. 
Activate/deactivate the high-temperature measuring range (
350.0 - 
550.0
 °C or 
662.0 - 1022
 °F), Measuring range page 25, after 
mounting or removing the high-temperature filter, otherwise the 
specified measuring accuracy is not ensured  
 
Mount tripod adapter 
Using the included tripod adapter, the imager can be fastened to 
the testo tripod (accessory) or to a commercially available standard 
tripod. With the tripod adapter mounted, the battery cannot be 
changed!  
1.
  Place the tripod adapter on the lower end of the handle and 
screw on using the included Allen key (ISO 2936, size 4).  
2.
  Slide thermal imager into the fastening plate of the testo tripod 
and lock or fasten it to the commercially available standard 
tripod (screw-type fastening). 
 
Mount sunshield 
The sunshield increases the readability of the display with a high 
exposure to light. 
>  Set the sunshield on the thermal imager from above (1) and pull 
both sides of the sunshield down around the display (2).