8 Tips and assistance 
 
 
39 
 
 
Tips and assistance 
 
8.1. 
Questions and answers 
 
Question Possible 
causes/solution 
Message 
Backup battery 
empty or not present
 is 
shown. 
1.
  Confirm the message with 
[OK]
2.
  Change buffer battery, see 
Change buffer battery page 37. 
Error 
No memory card 
inserted!
 is shown. 
Memory card defective or not inserted.
1.
  Confirm the message with 
[OK]
2.
  Check or insert memory card. 
Error 
Memory card full!
 is 
shown. 
Insufficient memory present. 
1.
  Confirm the message with 
[OK]
2.
  Insert new memory card. 
Error 
No lens/wrong lens!
is shown. 
1.
  Switch off thermal imager. 
2.
  Check lens. 
Message 
Charging not 
possible!
 is shown. 
Ambient temperature is outside the 
permissible range for charging the 
battery. 
1.
  Confirm the message with 
[OK]
2.
  Keep to the permissible ambient 
temperature. 
Error 
Permissible 
instrument temperature 
exceeded!
 is shown. 
1.
  Switch off thermal imager. 
2.
  Let instrument cool off and keep to 
the permissible ambient 
temperature. 
---
 or 
+++
 is shown instead 
of the reading. 
Reading is outside of the measuring 
range. 
>  Change measuring range. 
xxx
 is shown instead of the 
reading. 
Reading cannot be calculated. 
>  Check parameter settings for 
plausibility. 
 
If we could not answer your question, please contact your dealer or 
Testo Customer Service. For contact details see the rear side of 
this document or the web page www.testo.com/service-contact