Lecroy WaveJet 354A 500 MHz 4 Channel Oscilloscope WaveJet 354A Data Sheet
Product codes
WaveJet 354A
WaveJet 300A Series
10
WJ-A-GS-E Rev A
temporary conductivity caused by
condensation must be expected.
Protection Class 1 refers to a grounded
equipment, in which protection against
electric shock is achieved by Basic
Insulation and by means of a connection to
the protective ground conductor in the
building wiring.
condensation must be expected.
Protection Class 1 refers to a grounded
equipment, in which protection against
electric shock is achieved by Basic
Insulation and by means of a connection to
the protective ground conductor in the
building wiring.
Cooling Requirements
The instrument relies on forced air cooling
with internal fans and ventilation openings.
Care must be taken to avoid restricting the
airflow around the apertures (fan holes) at
each side of the scope. To ensure adequate
ventilation it is required to leave a10 cm (4
inch) minimum gap around the sides of the
instrument.
with internal fans and ventilation openings.
Care must be taken to avoid restricting the
airflow around the apertures (fan holes) at
each side of the scope. To ensure adequate
ventilation it is required to leave a10 cm (4
inch) minimum gap around the sides of the
instrument.
CAUTION
Do not block the ventilation holes
located on both sides of the scope.
located on both sides of the scope.
CAUTION
Do not allow any foreign matter to enter
the scope through the ventilation holes,
etc.
the scope through the ventilation holes,
etc.
AC Power Source
The instrument operates from a single-
phase, 100 to 240 V
phase, 100 to 240 V
rms
(+/-10%) AC power
source at 50/60 Hz (+/-5%), or single-phase
100 to 120 V
100 to 120 V
rms
(+/-10%) AC power source at
400 Hz (+/-5%).
No manual voltage selection is required
because the instrument automatically adapts
to line voltage.
because the instrument automatically adapts
to line voltage.
Depending on the accessories installed (PC
port plug-ins, Ethernet & GPIB options, etc.),
the instrument can draw up to 75 W (75 VA).
port plug-ins, Ethernet & GPIB options, etc.),
the instrument can draw up to 75 W (75 VA).
Note:
The instrument automatically adapts
itself to the AC line input within the
following ranges:
itself to the AC line input within the
following ranges:
Voltage
Range:
Range:
90 to 264 V
rms
90 to 132 V
rms
Frequency
Range:
Range:
47 to 63 Hz
380 to 420 Hz