Lecroy WaveJet 354A 500 MHz 4 Channel Oscilloscope WaveJet 354A Data Sheet

Product codes
WaveJet 354A
Page of 67
Getting Started Manual 
WJ-A-GS-E Rev A 
 
59
 
Saving and Recalling Waveforms 
Reference waveforms can be saved in internal memory (five locations) or in USB 
memory (limited by memory capacity of USB device). When you save a waveform, the 
setup is saved also.  
You can display up to five reference waveforms at the same time. Press REF to turn 
them all off together. 
The “Reference Waveform” menu is accessed by pressing the 
R
EF
 front panel button. 
 
 
 
 
 
 
Select from five memory locations (REF 1 to REF 5). To 
scroll through the choices, you can use the A
DJUST
 knob or 
press the Ref. No. button repeatedly. 
 
 
Select On to display the reference waveform. All reference 
waveforms appear white. 
 
 
When saving a reference waveform, select the source. 
 
 
When Recall REF Setup is selected, the setup stored in 
the memory location selected above (REF 1 to REF 5) will 
be recalled. This does not include setups saved in USB 
memory, which are recalled through S
AVE
/R
ECALL
 
When Save is selected, the source will be saved into the 
memory location selected above (REF 1 to REF 5).