According to modern ideas about danger of small doses of radiation,
the defensive measures of ngabited buildings should be conducted if a dose
rate of gamma radiation inside a building exceeds a dose rate outside for
more than
.
The estimation of a dose rate outside (background) is carried out near to
the inspected building not less than in 5 sighting points standing from 30 to
100 m apart from buildings and facilities and not closer than 20 m from each
other. The sighting points of estimation should be selected on location with a
natural soil without any man-caused modifications (road metal, sand,
asphalt) and radioactive contamination. The unit should be at 1 m high
above the earth surface.
The total time of estimation of a background dose rate will take 20 - 30
min es but if you want to receive authentic results it is necessary to conduct
a complete term examination.
For your convenience «Background» mode (page
) is supplemented
to the device which makes it easier to conduct examination.
0,20μSv/h
ut
25
Radiation control in quarters and public buildings.