Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Data Sheet

Product codes
BK-8500
Page of 76
Key
Display
I-set
If you press this key after powering up and there is no currently-
stored constant current value, you'll be prompted for the desired 
current level.  If there was a current value already stored, it will be 
momentarily displayed and the instrument will be in CC mode.  If 
you wish to change the current setting, press I-set again.  The 
OFF 
annunciator will be displayed.
On/Off
Turns the constant current load on.  The 
CC
 annunciator will be on.
 or 
Momentarily display the power level and the set current value.
On/Off
Turns the load off.
To operate the DC Load in CC mode in a transient condition, see the Transient operation section. 
To operate the DC Load in CC mode in a dynamic condition, see the Lists section.
Constant voltage mode
In constant voltage mode, the DC load will cause a constant voltage to appear at its terminals.  To 
set up the DC Load to operate in constant voltage mode and in the steady state condition, use the 
following keystrokes:
Key
Display
V-set
If you press this key after powering up and there is no currently-
stored constant voltage value, you'll be prompted for the desired 
voltage level.  If there was a voltage value already stored, it will be 
momentarily displayed and the instrument will be in CV mode.  If 
you wish to change the voltage setting, press V-set again.  The 
OFF
 annunciator will be displayed.
On/Off
Turns the constant voltage load on.  The 
CV
 annunciator will be 
on.
 or 
Momentarily display the power level and the set voltage value.
On/Off
Turns the load off.
To operate the DC Load in CV mode in a transient condition, see the Transient operation section. 
To operate the DC Load in CV mode in a dynamic condition, see the Lists section.
Constant power mode
In constant power mode, the DC load will cause a constant power to be dissipated in the load.  To 
set up the DC Load to operate in constant power mode and in the steady state condition, use the 
following keystrokes:
Key
Display
P-set
If you press this key after powering up and there is no currently-
stored constant power value, you'll be prompted for the desired 
8500 DC Load Series                    
Version:  September 2, 2009 
Page 30 of 76