Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Data Sheet
Product codes
BK-8500
Byte
offset
Meaning
3 is constant resistance (CR)
4-24
Reserved
0x3B Read the list operation (CC/CV/CW/CR)
Byte
offset
Meaning
3
List operation mode:
0 is constant current (CC)
1 is constant voltage (CV)
2 is constant power (CW)
3 is constant resistance (CR)
1 is constant voltage (CV)
2 is constant power (CW)
3 is constant resistance (CR)
4-24
Reserved
0x3C Set how lists repeat (ONCE or REPEAT)
Byte
offset
Meaning
3
How lists repeat:
0 is ONCE
1 is REPEAT
1 is REPEAT
4-24
Reserved
0x3D Read how lists repeat (ONCE or REPEAT)
Byte
offset
Meaning
3
How lists repeat:
0 is ONCE
1 is REPEAT
1 is REPEAT
4-24
Reserved
0x3E Set the number of list steps
Byte
offset
Meaning
3 to 4
2 byte little-endian integer for number of steps
5-24
Reserved
0x3F Read the number of list steps
Byte
offset
Meaning
3 to 4
2 byte little-endian integer for number of steps
5-24
Reserved