Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Data Sheet

Product codes
BK-8500
Page of 76
Byte 
offset
Meaning
3
Calibration protection state:
Bit 0:  value of 0 means the calibration state is not protected
Bit 0:  value of 1 means the calibration state is protected
4-24
Reserved
0x62   Set voltage calibration point index
Byte 
offset
Meaning
3
Voltage calibration point.  Valid values are 0x01 to 0x04.
4-24
Reserved
The DC Load has four calibration points for voltage.  These would typically span the full range of the 
instrument's voltage range.
0x63   Send the actual voltage to the calibration program
Byte 
offset
Meaning
3 to 6
4 byte little-endian integer representing the voltage in units of mV
7-24
Reserved
0x64   Set current calibration point index
Byte 
offset
Meaning
3
Current calibration point.  Valid values are 0x01 to 0x04.
4-24
Reserved
The DC Load has four calibration points for current.  These would typically span the full range of the 
instrument's current range.
0x65   Send the actual current to the calibration program
Byte 
offset
Meaning
3 to 6
4 byte little-endian integer representing the current in units of 0.1 mA
7-24
Reserved
0x66   Store the calibration data to EEPROM
Byte 
offset
Meaning
3-24
Reserved
8500 DC Load Series                    
Version:  September 2, 2009 
Page 68 of 76