Gw Instek PEL-3111 Electronic Load PEL-3111 Data Sheet

Product codes
PEL-3111
Page of 174
 
 
COMMNAND OVERVIEW 
91 
Power Subsystem Commands 
 
 
:POWer[:VA] ......................................................................... 91 
:POWer:VB............................................................................ 92 
:POWer:L1 ............................................................................. 92 
:POWer:L2 ............................................................................. 93 
:POWer:SET ......................................................................... 93 
:POWer:LEVel ...................................................................... 94 
:POWer:T1 ............................................................................ 95 
:POWer:T2 ............................................................................ 95 
:POWer:FREQuency ........................................................... 96 
:POWer:DUTY ..................................................................... 97
 
 
:POWer[:VA] 
Set
Query
 
 
Description 
Sets or queries the CP mode “A Value” power. 
This command is applicable to static mode only. 
Note: A different “A Value” can be set for different 
current ranges.   
Note: The [:VA]  node can only be omitted when in 
static mode. 
Syntax 
:POWer[:VA] {<NRf>[ W ] | MINimum | MAXimum } 
Query Syntax 
:POWer[:VA]? 
Parameter  
<NRf>[ W ] 
“A Value” power value 
 
MINIMUM or MIN 
Minimum power level 
 
MAXIMUM or MAX 
Maximum power level 
Return parameter  <NR2> 
Returns the “A Value” 
power value. 
Example 
:POWer:VA MIN 
Sets the power value to the minimum. 
Query example 
:POWer:VA? 
>10W 
Returns the “A Value” power setting.