Gw Instek PEL-3211 Electronic Load PEL-3211 Data Sheet

Product codes
PEL-3211
Page of 174
 
 
COMMNAND OVERVIEW 
151 
Status Registers  
To program the PEL-3000 Series effectively, the Status Register 
structure needs to be understood. This chapter explains in detail 
the structure of the status registers. 
 
Overview  
The status registers are used to determine the 
status of the electronic load. The status registers 
maintain the status of the protection conditions, 
load conditions and channel conditions of the 
load modules. 
The PEL-3000 series have a number of register 
groups: 
CSummary Registers 
Questionable Status Registers 
Standard Event Status Registers 
Status Byte Register 
Service Request Generation Register 
Each register group consists of a Condition, 
EVENT and ENABLE register. They also have 
PTR and NTR (positive and negative transition) 
filters. 
The structure of the status registers is shown on 
the next page.