PEL-3000 Programming Manual 
6
 
Configure GPIB Interface  
To use GPIB, the optional GPIB port must be installed. See the user 
manual for details. 
 
Operation 
1.  Ensure the PEL-3000 is off before proceeding. 
 
2.  Connect a GPIB cable from a GPIB controller to 
the GPIB port on the PEL-3000. 
3.  Turn the PEL-3000 on. 
4.  Press 
Shift
 + 
Utility
Help
 > Interface[F3] and set 
the Interface setting to GPIB
 
5.  Set the GPIB address. 
 
GPIB address 
0~30 
GPIB constraints 
 
Maximum 15 devices altogether, 20m cable length, 
2m between each device 
 
Unique address assigned to each device 
 
At least 2/3 of the devices turned On 
 
No loop or parallel connection 
 Pin Assignment 
1
12
13
24
 
 
Pin  Signal 
Pin 
Signal 
 
1~4  Data I/O 1~4 
13~16  Data I/O 5~8 
 
EOI 
17 
REN 
 
DAV 
18 
Ground (DAV) 
 
NRFD 
19 
Ground (NRFD) 
 
NDAC 
20 
Ground (NDAC) 
 
IFC 
21 
Ground (IFC) 
 
10  SRQ 
22 
Ground (SRQ)