Gw Instek PEL-3041 Electronic Load PEL-3041 Data Sheet

Product codes
PEL-3041
Page of 119
 
 
PEL-3000 User Manual 
68
 
Display 
Mode
CC
I Range
H 35A
V Range
H 150V
Function
Dynamic
Configure
0.00
V
0.000
A
0.00
w
Timer1
0.025
SlewRate
2500.00
SlewRate
2500.00
LOAD
RS232
03/Sep/2012
ms
mA/uS
mA/uS
Slewrate settings
Slewrate settings
 
CV Mode Response Speed 
 
Description 
The response speed setting is the response 
speed for the negative feedback control of the 
load current when used in CV mode. Response 
speed settings are only applicable to CV mode. 
 
A response speed that is too fast could cause the 
unit to be unstable.  
 
Reducing the response speed can improve 
stability.  
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
. Make sure the unit is in CV 
mode by using the Mode[F1] soft-key. 
 
3.  Select the response speed with the Response[F4] 
soft-key.  
 
Range:  
Slow, Fast