8.  Choose the P/F-Time (pass/fail time) in 
seconds. 
 
The P/F-Time refers to the P/F delay time. This 
delay time includes the 0.06 P/F start test time, 
as shown in the timing diagram on page 94.  
 
P/F-Time 
Off, 0.0 ~ 119.9 seconds 
 
9.  Set the Short-Time in seconds. 
 
Has the same action as pressing the short key. 
See page 61 for details about shorting the load. 
 
Short-Time 
Off, 0.1 seconds ~ On-Time 
 
10. Repeat steps 3 to 9 for all the steps in the 
program. 
 
A maximum of 16 steps per program can be 
created. 
 
Steps that are not configured are set to ―Skip‖ 
by default. 
 
11. Press Save[F3] to save the program and all the 
steps in the program. 
 
The program will be saved to internal memory.  
 
See the Save/Recall chapter on details on how 
to save to Setup memory. 
Recall Default 
Pressing Recall Default[F4] will recall the default 
settings for each program/step. See page 178 
for details.