OPERATION 
55 
CR Units  
 
Description 
The CR setting units can be set to ohm (Ω) or 
millisiemens (mS).  
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
 > Configure[F5] > Other[F2] and 
set the CR Unit setting. 
 
Range:  
Ω, mS 
CV Mode  
 
Description 
In Constant Voltage Mode, the unit will 
maintain a constant voltage. In CV mode you 
set the constant voltage level. For more details 
on CV mode, see the appendix on page 190. 
 Warning 
If you change the mode or the range when the load 
is already on, the load will be turned off 
automatically.  
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
3.  Select CV mode with the Mode[F1] soft-key.  
4.  Select the current range with the I Range[F2] 
soft-key.  
 
Range:  
High, Middle, Low 
 
5.  Select the voltage range with the V Range[F3] 
soft-key.  
 
Range:  
High, Low