OPERATION 
73 
Von Voltage Delay 
 
Description 
Von Delay is the amount of time the unit will 
wait before turning the load on after the Von 
Voltage threshold has been latched. This will 
prevent overshoot current from affecting the 
Von Voltage threshold. 
Operation 
2.  Press 
Main
 > Configure[F5] > Other[F2] and 
set the Von Delay time.  
 
Range:  
Von Delay:  OFF, 1.0~60ms 
 
 
Note: CR mode can have the delay time set 
separately from the other modes (called 
Von Delay 
–CR
 when in CR mode). 
Timer Functions 
Count Time 
 
Description 
When Count Time is set to on, it will count the 
elapsed time from when the load was turned on 
to when it was turned off.   
 
This function is applicable to manual and 
automatic shutdown (such as from protection 
functions such as UVP etc.) 
 
The elapsed time will be shown in the display 
Measurement area.