Triggering on Waveforms 
 
Triggering on Pulses 
4
 
  
63 
Tip 
The test tool stores all triggered screens in the 
replay memory. For example, if you setup your 
triggering for glitches, you can capture 100 
glitches with time stamps. Use the 
REPLAY
 key to 
look at all the stored glitches. 
 
Figure 28. Triggering on Narrow Glitches 
Finding Missing Pulses 
The next example covers finding missing pulses in a train 
of positive pulses. In this example it is assumed that the 
pulses have a 100 ms distance between the rising edges. 
If the time accidently increases to 200 ms, a pulse is 
missing. To set the test tool to trigger on such missing 
pulses, let it trigger on gaps bigger than about 110 ms. 
Do the following: 
TRIGGER
 
Display the 
TRIGGER 
key labels. 
 
F4
 
Open the 
TRIGGER OPTIONS 
menu. 
 
ENTER
Select Pulse Width on A... to 
open the 
TRIGGER ON PULSE WIDTH 
menu.