  Warning 
To avoid electrical shock or fire: 
• Use only the Fluke power supply, Model 
BC190 (Power Adapter). 
• Before use check that the selected/indicated  
range on the BC190 matches the local line 
power voltage and frequency. 
• For the BC190/808 universal Power Adapter 
only use line cords that comply with the local 
safety regulations. 
Note: 
To accommodate connection to various line 
power sockets, the BC190/808 universal Power 
Adapter is equipped with a male plug that must 
be connected to a line cord appropriate for local 
use. Since the adapter is isolated, the line cord 
does not need to be equipped with a terminal for 
connection to protective ground.  Since line cords 
with a protective grounding terminal are more 
commonly available you might consider using 
these anyhow. 
 Warning 
To avoid electrical shock or fire if a product 
input is connected to more than 42 V peak 
(30 Vrms) or 60 V dc: 
• Use only insulated voltage probes, test leads 
and adapters supplied with the product, or 
indicated by Fluke as suitable for the Fluke 
190 Series II ScopeMeter series. 
• Before use, inspect voltage probes, test leads 
and accessories for mechanical damage and 
replace when damaged. 
• Remove all probes, test leads and 
accessories that are not in use. 
• Always connect the power adapter first to the 
ac outlet before connecting it to the product. 
• Do not touch voltages >30 V ac rms, 42 V ac 
peak, or 60 V dc. 
• Do not connect the ground spring (figure 1, 
item d) to voltages higher than 42 V peak 
(30 Vrms) from earth ground. 
• Do not apply more than the rated voltage, 
between the terminals or between each 
terminal and earth ground.