Kern Counting scales CXB 15K1 Weight range 15 kg Readability 1 g mains-powered, rechargeable Silver CXB 15K1 User Manual

Product codes
CXB 15K1
Page of 54
 
CXB/CXP-BA-e-1016 
20 
7.6  Adjustment with external weight 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated -  in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To re-
ceive accurate measuring values it is also recommended to adjust the balance peri-
odically in weighing operation.  
Procedure when adjusting: 
In calibrated balances the adjustment function is switch locked.  
To carry out calibrations, you will have to throw the unlock switch 
Observe stable environmental conditions. A warming up time (see chapter 1) is re-
quired for stabilization. Ensure that there are no objects on the weighing plate. 
 
7.6.1  Adjustment of models CXP 
 
Operation 
Ensure that there are no objects on the weighing plate. 
0
W
a
PRE-
TARE
 
Press the 
 key and keep it pressed down whilst simultaneously operating the 
 
key  
Weight value of adjustment weight is flashing on display. However, you may enter 
a value of your choice via the number keys. * 
3000
W
a
PRE-
TARE
 
CAL
 
Carefully place adjusting weight in the centre of the weighing plate. Adjustment will 
be carried out automatically after dead stop control.  
Remove adjustment weight during count down to zero. 
The balance returns automatically into weighing mode. An error message will ap-
pear on the display should a adjustment error occur or should the adjustment 
weight be incorrect. Turn balance off, then restart it and repeat the adjustment 
process.