7.6.3  Adjustment of models CXB_M 
 
 
Operation 
Ensure that there are no objects on the weighing plate. 
Ö
 Press 
 + 1 + 5 keys 
Ö
 Press 
 key 
The display shows: 
0.000
 
M+
0
 
12345
 
Ö
 Press   key 
The display shows: 
15.000
 
(Example) 
M+
0
 
12345
 
Weight value of adjustment weight is flashing on display. 
Ö
  Carefully place adjusting weight in the centre of the weighing plate. 
Ö
 Press   key 
Adjustment will be carried out automatically after dead stop control.  
Ö
  Remove adjustment weight during count down to zero. 
The balance returns automatically into weighing mode. An error message will ap-
pear on the display should a adjustment error occur or should the adjustment 
weight be incorrect. Turn balance off, then restart it and repeat the adjustment 
process. 
 
* The adjustment should be made with the recommended adjustment weight (see 
chap. 1 "Technical data"). Weights of different nominal values may be used for ad-
justment but are not optimal for technical measuring.  
Info about adjustment weights can be found on the Internet at: http://www.kern-
sohn.com