Kern Counting scales CPB 15K0.2N Weight range 15 kg Readability 0.2 g mains-powered, rechargeable Silver CPB 15K0.2N User Manual

Product codes
CPB 15K0.2N
Page of 49
CPB-N / CPB-DM-BA-e-1323 
23 
6.9.1  Adjustment switch and seals  
Possible seals: 
 enforced, and 
 or 
 
 
 
 
1.  Seal mark 1 
2.  Cover 
3.  Verification switch 
4.  Verification wire 
 
6.10  Checking the balance verification settings  
For the adjustment, the balance must be switched over to service mode. 
 
 
In the service mode the parameters of the balance can be modified.  
The service parameters may not be modified, as this could damage the 
balance settings. 
 
In calibrated scales the service mode is locked individually for each switch. To 
disable the access lock, destroy the seal and actuate the switch. For position of 
switch see chap. 6.9.1. 
Attention
After destruction of the seal the balance must be re-verified by an authorised agency 
and a new verification wire/seal mark fitted before it can be reused for applications 
subject to verification.