6.2.2 Scope of delivery
Serial accessories:
Balance
Weighing plate
Power cable
Protective cover
Instruction Manual
6.3 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage.
Only use original KERN mains adapters. Using other makes requires consent by
KERN.
the same as the local voltage.
Only use original KERN mains adapters. Using other makes requires consent by
KERN.
6.4 Rechargeable battery operation (optional)
The internal battery is charged with the supplied mains cable.
Before the first use, the rechargeable battery should be charged by connecting it to
the mains power cable for at least 12 hours. The operating time of the battery is
about. 70h. Charging time until complete recharging ca. 12h.
AUTO-OFF function can be selected after 3, 5, 15, 30 min. to save the rechargeable
battery (see chap.12).
Before the first use, the rechargeable battery should be charged by connecting it to
the mains power cable for at least 12 hours. The operating time of the battery is
about. 70h. Charging time until complete recharging ca. 12h.
AUTO-OFF function can be selected after 3, 5, 15, 30 min. to save the rechargeable
battery (see chap.12).
If an arrow appears on the weight display
above the battery symbol
or "bat lo" when turning on the balance, this is an indication that the capacity of the
rechargeable battery will soon be exhausted. The balance will be ready to operate
for about another 10 hours, then it will switch off automatically. Connect the power
cable as soon as possible to load the rechargeable battery.
The LED display under the piece number window informs you during charging about
the charging status of the rechargeable battery.
rechargeable battery will soon be exhausted. The balance will be ready to operate
for about another 10 hours, then it will switch off automatically. Connect the power
cable as soon as possible to load the rechargeable battery.
The LED display under the piece number window informs you during charging about
the charging status of the rechargeable battery.
red:
Battery is almost discharged
green: Battery is completely discharged
6.5 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
6.6 Initial Commissioning
A warming up time of 2 hours after switching on stabilizes the measuring values.
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.