9  Adjustment 
 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated -  in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To 
receive accurate measuring values it is also recommended to adjust the balance 
periodically in weighing operation.  
 
 
  For verified balances the adjustment is locked. In order to access the 
menu you will have to short-circuit the two contacts of the circuit board 
with a jumper (See chap. 9.3). 
Attention: 
After destruction of the seal the balance must be re-verified by an 
authorised agency and a new verification wire/seal mark fitted before it 
can be reused for applications subject to verification. 
Observe the verification notes (see chapter 9.3). 
  The weight to be used depends on the capacity of the balance. Carry 
out adjustment as near as possible to the balance’s maximum weight.
 
Info about test weights can be found on the Internet at: http://www.kern-
  Observe stable environmental conditions.  
Stabilisation requires a certain warm-up time.