Kern Counting scales CXB 3K0.2 Weight range 3 kg Readability 0.2 g mains-powered, rechargeable Silver CXB 3K0.2 User Manual

Product codes
CXB 3K0.2
Page of 45
CXB/CXP-BA-e-0711 
21
8.2  Numeric entering of the reference weight  
 
If you know the reference weight/piece you can enter this via number keys. 
 
Enter reference weight via number keys   
W
a
PRE-
TARE
 
1.833
 
0
M+
W
 
Confirm with 
 key 
W
a
PRE-
TARE
 
1.833
Reference weight 
0
M+
W
 
Now you can place the parts to be counted onto the weighing plate. All quantity pa-
rameters of your goods to be weighed are displayed: 
 
 
8.3  Automatic reference optimization 
If it was impossible to determine a reference due to instable goods to be weighed or 
an insufficient reference weight, the [
W ] display will appear in the reference weight 
window during reference calculation. 
 
Overlay 
W indicates: 
 
Placed number of pieces insufficient for reference calculation 
CXB models < 40 d 
CXP models < 20 d 
 
Placed reference weight insufficient for reference calculation 
CXB models < 4/5 d 
CXP models < 1/5 d 
 
Add additional parts until the [
W ] display disappears. 
An audio signal indicates that reference optimization has been carried out.  
At every reference optimisation, the reference weight is calculated anew. As the addi-
tional pieces increase the base for the calculation, the reference also becomes more 
exact.