Kern Counting scales CXB 3K0.2 Weight range 3 kg Readability 0.2 g mains-powered, rechargeable Silver CXB 3K0.2 User Manual

Product codes
CXB 3K0.2
Page of 45
CXB/CXP-BA-e-0711 
41
15  Instant help  
In case of an error in the program process, briefly turn off the balance and disconnect 
from power supply. The weighing process must then be restarted from the beginning. 
 
Fault 
Possible cause 
•  The balance is not switched on. 
•  The mains supply connection has been 
interrupted (mains cable not plugged 
in/faulty). 
•  Power supply interrupted. 
•  Batteries are inserted incorrectly or empty
The displayed weight does not glow. 
•  No batteries inserted. 
•  Draught/air movement 
•  Table/floor vibrations 
The displayed weight is permanently 
changing 
•  Weighing plate has contact with other 
objects. 
•  Electromagnetic fields / static charging 
(choose different location/switch off inter-
fering device if possible) 
•  The display of the balance is not at zero 
•  Adjustment is no longer correct. 
The weighing result is obviously in-
correct 
•  Great fluctuations in temperature. 
Electromagnetic fields / static charging 
(choose different location/switch off inter-
fering device if possible) 
 
Should other error messages occur, switch balance off and then on again. If the error 
message remains inform manufacturer.