Fluke Calibrator, 4353499 Data Sheet
Product codes
4353499
2 Fluke Corporation Fluke 721 Precision Pressure Calibrator
Low pressure sensor
High pressure sensor
Model
Description
Range sensor 1
Resolution
sensor 1
Accuracy
sensor 1
Range sensor 2
Resolution
sensor 2
Accuracy
sensor 2
Fluke-721-1601 Dual Sensor Pressure Cali-
brator, 16 PSIG, 100 PSIG
-14 psi to +16 psi
-0.97 bar to 1.1 bar
-0.97 bar to 1.1 bar
0.001 psi,
0.0001 bar
0.0001 bar
0.025 % of
full scale
full scale
-12 psi to +100 psi
-0.83 bar to 6.9 bar
-0.83 bar to 6.9 bar
0.01 psi
0.0001 bar
0.0001 bar
0.025 % of
full scale
full scale
Fluke-721-1603 Dual Sensor Pressure Cali-
brator, 16 PSIG, 300 PSIG
-12 psi to +300 psi
-0.83 bar to 20 bar
-0.83 bar to 20 bar
0.01 psi
0.001 bar
0.001 bar
Fluke-721-1605 Dual Sensor Pressure Cali-
brator, 16 PSIG, 500 PSIG
-12 psi to +500 psi
-0.83 bar to 34.5 bar
-0.83 bar to 34.5 bar
0.01 psi
0.001 bar
0.001 bar
Fluke-721-1610 Dual Sensor Pressure Cali-
brator, 16 PSIG, 1000 PSIG
0 psi to +1000 psi
0.00 bar to 69 bar
0.00 bar to 69 bar
0.1 psi
0.001 bar
0.001 bar
Fluke-721-1615 Dual Sensor Pressure Cali-
brator, 16 PSIG, 1500 PSIG
0 psi to +1500 psi
0.00 bar to 103.4 bar
0.00 bar to 103.4 bar
0.1 psi
0.001 bar
0.001 bar
Fluke-721-1630 Dual Sensor Pressure Cali-
brator, 16 PSIG, 3000 PSIG
0 psi to +3000 psi
0.00 bar to 200 bar
0.00 bar to 200 bar
0.1 psi
0.01 bar
0.01 bar
Fluke-721-1650 Dual Sensor Pressure Cali-
brator, 16 PSIG, 5000 PSIG
0 psi to +5000 psi
0.00 bar to 345 bar
0.00 bar to 345 bar
0.1 psi
0.01 bar
0.01 bar
0.035 % of
full scale
full scale
Fluke-721-3601 Dual Sensor Pressure Cali-
brator, 36 PSIG, 100 PSIG
-14 psi to +36 psi
-0.97 bar to 2.48 bar
-0.97 bar to 2.48 bar
-12 to +100 psi
-0.83 bar to 6.9 bar
-0.83 bar to 6.9 bar
.01 psi
0.0001 bar
0.0001 bar
0.025 % of
full scale
full scale
Fluke-721-3603 Dual Sensor Pressure Cali-
brator, 36 PSIG, 300 PSIG
-12 psi to +300 psi
-0.83 bar to 20 bar
-0.83 bar to 20 bar
0.01 psi
0.001 bar
0.001 bar
Fluke-721-3605 Dual Sensor Pressure Cali-
brator, 36 PSIG, 500 PSIG
-12 psi to +500 psi
-0.83 bar to 34.5 bar
-0.83 bar to 34.5 bar
0.01 psi
0.001 bar
0.001 bar
Fluke-721-3610 Dual Sensor Pressure Cali-
brator, 36 PSIG, 1000 PSIG
0 psi to +1000 psi
0.00 bar to 69 bar
0.00 bar to 69 bar
0.1 psi
0.001 bar
0.001 bar
Fluke-721-3615 Dual Sensor Pressure Cali-
brator, 36 PSIG, 1500 PSIG
0 psi to +1500 psi
0.00 bar to 103.4 bar
0.00 bar to 103.4 bar
0.1 psi
0.01 bar
0.01 bar
Fluke-721-3630 Dual Sensor Pressure Cali-
brator, 36 PSIG, 3000 PSIG
0 psi to +3000 psi
0.00 bar to 200 bar
0.00 bar to 200 bar
0.1 psi
0.01 bar
0.01 bar
Fluke-721-3650 Dual Sensor Pressure Cali-
brator, 36 PSIG, 5000 PSIG
0 psi to +5000 psi
0.00 bar to 345 bar
0.00 bar to 345 bar
0.1 psi
0.01 bar
0.01 bar
0.035 % of
full scale
full scale
Temperature effect
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C
General specifications
Instrument setup recall
Instrument setup recall
Last used on power-up
Environmental
Operating temperature -10 °C
to +50 °C
Storage temperature (without
batteries) -20 °C to +60 °C
to +50 °C
Storage temperature (without
batteries) -20 °C to +60 °C
Altitude
2,000 meters
Ingress protection
IP-40
Power requirements
6.0 v dc
Battery life
>35 hours, typical usage
Batteries required
(4) Standard AA cells
Dimentions (HxWxD)
7.9” X 4.3” X 2.3” (20 cm X
11 cm X 5.8 cm)
11 cm X 5.8 cm)
Weight
03539 kg (1 lb 3 oz)
Connectors/ports pressure Two, 1/8 in NPT, 700P Series
Pressure Module; RTD Probe
Specifications 1-year
Electrical and temperature measurement 1-year
Function
Function
Range
Resolution
Accuracy
mA measure
0 to 24 mA
0.001 mA
±0.015% of rdg
±0.002mA
±0.002mA
VDC measure
0 to 30 VDC
0.001V
±0.015% of rdg
±0.002V
±0.002V
*Temperature
measure (RTD/
Ohms)
measure (RTD/
Ohms)
-40 °C to 150 °C
(-40 °F to 302 °F)
(-40 °F to 302 °F)
0.01 °C, 0.01 °F
±0.015% of rdg
±0.02 ohms;
±0.1°C (±0.2°F)
±0.25 degC (±0.45
degF) combined
uncertainty when
using 720 RTD
probe accessory
±0.02 ohms;
±0.1°C (±0.2°F)
±0.25 degC (±0.45
degF) combined
uncertainty when
using 720 RTD
probe accessory
Loop power supply 24V
N/A
24 mA @ 24V
*Temperature measurement requires optional Fluke-720RTD Pt-100 RTD probe
available as an accessory