Fluke Calibrator, 4353229 Data Sheet

Product codes
4353229
Page of 80
 
Pressure Calibrator 
 
Product Features 
13 
Zero Function Use 
If a pressure mode is selected and pressure is within the 
zero limit, the Product zeros the active display when Z is 
pushed. The zero limits are within 10 % of the full-scale 
range of the selected sensor. If the display shows “OL,” 
the zero function will not operate. 
Note 
Z
 is only used for pressure. 
Internal Sensor and Pressure Module (non-
absolute) 
When a sensor or module is selected on the active 
display, and Z is pushed, the Product subtracts the 
current reading from the output. The zero limits are within 
10 % of the full-scale range of the selected sensor. If the 
display shows “OL,” the zero function will not operate. 
Absolute Pressure 
When an absolute pressure range is selected on the 
active display, and Z is pushed, the Product prompts for 
the barometric reference pressure as in the screen below.  
 
hix023.eps 
Push  and  to use the on-screen arrows to record the 
correct barometric reference pressure. The sensor port 
should be open (vented) to atmosphere while this 
procedure is done.