24 
DAMP 
Turn on or off Damping with the DAMP menu selection. 
When damping is on, the Product shows a running 
average from ten measurements. The Product makes 
approximately three indications per second.  
To use the Damping function: 
1. Push 
 to access the menus. 
2. Push 
 to step through the menu until DAMP is 
shown on the menu bar.  
3. Push 
 to select the DAMP menu.  
4. Push 
 or  to turn on or off the DAMP function. 
5. Push 
 when finished. 
 
hmq064.eps 
Measure Pressure 
To measure pressure, connect the Product with the 
correct fitting and select a pressure port. The Product has 
two internal sensors and many optional external sensors 
(EPMs) are available. Make sure to choose the sensor 
based on working pressures and accuracy. 
Warning 
To prevent personal injury: 
 
Pressure sensors can be damaged 
and/or personnel injury can occur due to 
improper application of pressure. Refer 
to Table 4 for information on 
overpressure and burst pressure ratings. 
Vacuum should not be applied to any 
gauge pressure sensor. The Product 
display shows “OL” when an 
inappropriate pressure is applied. If “OL” 
is shown on any pressure display, the 
pressure should be reduced or vented 
immediately to prevent Product damage 
or possible personnel injury. “OL” is 
shown when the pressure exceeds 110 % 
of the nominal range of the sensor or 
when a vacuum in excess of 2 PSI is 
applied on gauge range sensors.