Fluke Calibrator, 4353366 Data Sheet

Product codes
4353366
Page of 48
721 
Users Manual 
10 
Home Menu Functionality 
There are three options for the Home Menu:  
 P1/P2 
 mA/V/RTD 
 MENU 
These options are shown across the bottom of the 
display. 
From anywhere within the menu structure, push  to get 
back to the Home Menu. 
Backlight Use 
Push  to toggle on and off the backlight. This function 
cannot be controlled by the serial interface. 
Zero Function 
In pressure mode and when the pressure is within the 
zero limit, The Product zeros pressure on any port that is 
currently shown on the display. The zero limits are within 
10 % of the full-scale range of the selected sensor. If the 
display shows “OL,” the zero function will not operate. 
External Sensor (Absolute) 
If the active display shows a selected absolute module, 
and Z is pushed, push  (SET) and you are prompted 
to set or default the pressure reference. This is done with 
 and  (the up and down arrows). The sensor port 
must be open (vented) to atmosphere while this 
procedure is done. Push   (SET REF DONE) when 
zero adjust is complete.