Pressure Calibrator 
 
Media Compatibility 
25 
 
Push Z to zero the pressure sensor 
when vented to atmospheric pressure. 
Note 
 To ensure accuracy of the Product, the Product 
must be zeroed before a device is calibrated. 
See the “Zero Function Use” section. 
Media Compatibility 
The Product has a media-isolated sensor to prevent 
sensor contamination. Whenever possible, clean dry air is 
the media of choice. If this is not possible, make sure the 
media is compatible with Nickel Plated Brass and 316 
Stainless Steel.  
Measurements 
Use the input terminals on the front of the Product to 
measure current and volts. Current is measured in mA 
and percentage of range. The range on the Product is set 
at 0 % at 4 mA and 100 % at 20 mA. 
Use the RTD connector and an RTD probe to measure 
temperature. 
From the main menu, push  to select mA, Volts, or 
RTD. This function will only work on the LOWER screen. 
Note 
The display shows “OL” when the measured 
current is more than the nominal range of 
current measurement (24 mA). 
The display shows “OL” when the measured 
voltage is more than the nominal range of the 
voltage measurement (30 V).