Fluke Calibrator, 4353366 Data Sheet
Product codes
4353366
Pressure Calibrator
Specifications
35
Specifications
(15
°
C to 35
°
C unless otherwise noted)
Environmental
Operating Temperature ............................ -10
°
C to +50
°
C (14
°
F to +122
°
F)
Storage
With Batteries ........................................ Per battery manufacturer’s specification, not to exceed storage specification without batteries.
Without Batteries ................................... -20
°
C to +60
°
C (-4
°
F to +140
°
F)
Altitude ...................................................... 2000 m
Power Requirements ................................ 6 V dc
Batteries .................................................... 4 AA batteries (alkaline)
Battery Life ................................................ >35 hours, typical usage
Batteries .................................................... 4 AA batteries (alkaline)
Battery Life ................................................ >35 hours, typical usage
Electrical and Temperature Measurement (1 year)
Function Range
Resolution
Accuracy
mA Measure
0 to 24 mA
0.001 mA
±
0.015 % of rdg
±
0.002 mA
V dc Measure
0 to 30 V dc
0.001 V
±
0.015 % of rdg
±
0.002 V
*Temperature measure
(RTD/Ohms)
(RTD/Ohms)
-40
°
C to 150
°
C (-40
°
F to 302
°
F) 0.01
°
C, 0.01
°
F
±
0.015 % of rdg
±
0.02
Ω
;
±
0.1
°
C (
±
0.2
°
F)
±
0.25
°
C (
±
0.45
°
F) combined uncertainty when using
720 RTD probe accessory
Loop Power Supply
24 V
N/A
24 mA @ 24 V