Fluke Calibrator, 4353412 Data Sheet

Product codes
4353412
Page of 48
 
Pressure Calibrator 
 
Specifications 
35 
Specifications 
(15 
°
C to 35 
°
C unless otherwise noted) 
Environmental 
Operating Temperature  ............................ -10 
°
C to +50 
°
C (14 
°
F to +122 
°
F) 
Storage  
With Batteries ........................................ Per battery manufacturer’s specification, not to exceed storage specification without batteries. 
Without Batteries ................................... -20 
°
C to +60 
°
C (-4 
°
F to +140 
°
F) 
Altitude ...................................................... 2000 m 
Power Requirements  ................................ 6 V dc 
Batteries .................................................... 4 AA batteries (alkaline) 
Battery Life ................................................ >35 hours, typical usage 
Electrical and Temperature Measurement (1 year) 
Function Range 
Resolution 
Accuracy 
mA Measure 
0 to 24 mA 
0.001 mA 
±
0.015 % of rdg 
±
0.002 mA 
V dc Measure 
0 to 30 V dc 
0.001 V 
±
0.015 % of rdg 
±
0.002 V 
*Temperature measure 
(RTD/Ohms) 
-40 
°
C to 150 
°
C (-40 
°
F to 302 
°
F) 0.01 
°
C, 0.01 
°
±
0.015 % of rdg 
±
0.02 
Ω
±
0.1 
°
C (
±
0.2 
°
F) 
±
0.25 
°
C (
±
 0.45 
°
F) combined uncertainty when using 
720 RTD probe accessory 
Loop Power Supply 
24 V 
N/A 
24 mA @ 24 V 
*Temperature measurement requires optional 720RTD Pt-100 RTD probe available as an accessory.