Transmitter Calibration 
mA Input Function 
The mA input function reads back the 4 mA to 20 mA 
output from the device being calibrated. This can be done 
passively. The device under test directly generates 4 mA 
to 20 mA and can be read by the Product. 
 
hmq047.eps 
Pressure-to-Current Transmitter Calibration 
To calibrate a pressure-to-current transmitter (P/I): 
1.  Connect the Product and the pump to the transmitter. 
See Figure 8. 
2.  Apply pressure with the pump. 
3.  Measure the current output of the transmitter. 
4.  Ensure the reading is correct. If not, adjust the 
transmitter as necessary. 
Note 
Use low-volume tubing when possible.