Pressure Switch Test 
Connect a pressure switch to the Product as shown in 
Figure 8.  
Pressure
Switch
Use low volume tubing when possible
 
hix017.eps 
Figure 8. Pressure Switch Connection 
To do a pressure switch test: 
1.  Change the setup to Setup 4 (default switch test) 
listed under “Setups” in the MORE configuration 
menu choices. The upper display is set to [P1] ST, all 
other displays are off. 
Note  
The pressure switch test can be done with these 
functions: [P1] ST, or EXT ST. 
Use low-volume tubing when possible. 
2.  Connect the Product to the switch with the pressure 
switch terminals to the pressure switch contacts (de-
energized dry contacts). The polarity of the terminals 
does not matter. Connect the pump from the Product 
to the input of the pressure switch.  
3.  Make sure the vent on the pump is open.  
4.  Zero the Product if necessary.  
5.  Close the vent after the Product is zeroed. If a 
normally closed switch, the top of the display shows 
“CLOSE”. 
6.  Apply pressure with the pump slowly until the switch 
opens.