Fluke Calibrator, 4353283 Data Sheet

Product codes
4353283
Page of 48
721 
Users Manual 
30 
Ranges and Resolution 
Ranges and resolutions for the Product are shown in Table 4.
Table 4. Ranges and Resolutions 
Range 
(PSI)  
 
16  36 
100 300 500 1000 1500 3000 5000 
Burst 
Pressure 
 
 
60  120 
400  1200 2000 4000 6000 9000 10000 
Proof Pressure 
(PSI) 
 
 
35  70 
200  600  1000 2000 3000 6000 7000 
Engineering Unit 
Factor 
  
  
  
  
  
  
  
  
  
psi 
16  36 
100 300 500 1000 1500 3000 5000 
bar 
0.06894757  1.1032 2.4821  6.8947 20.684 34.474 68.947 103.42 206.84 344.74 
mbar 
68.94757 
1103.2 2482.1  6894.8 20684 34474 68948 N/A 
N/A 
N/A 
kPa 
6.894757 
110.32 248.21  689.48 2068.4 3447.4 6894.8 10342  20684  34474 
MPa 
0.00689476  0.1103 0.2482  0.6894 2.0684 3.4474 6.8948 10.342 20.684 34.474 
kg/cm2 
0.07030697  1.1249 2.5311  7.0307 21.092 35.153 70.307 105.46 210.92 351.53 
cmH20 @ 4 
°
70.3089 
1124.9 2531.1  7030.9 21093 35154 70309 N/A 
N/A 
N/A 
cmH20 @ 20 
°
70.4336 
1126.9 2535.6  7043.4 21130 35217 70434 N/A 
N/A 
N/A 
mmH20 @ 4 
°
703.089  11249 
25311 70309 
N/A N/A N/A N/A N/A N/A 
mmH20 @ 20 
°
704.336  11269 
25356 70434 
N/A N/A N/A N/A N/A N/A 
inH20 @ 4 
°
27.68067 
442.89 996.50  2768.1 8304.2 13840 27681 41521 83042 N/A 
inH20 @ 20 
°
27.72977 
443.68 998.27  2773.0 8318.9 13865 27730 41595 83189 N/A 
inH20 @ 60 
°
27.70759 
443.32 997.47  2770.8 8312.3 13854 27708 41561 83123 N/A 
mmHg @ 0 
°
51.71508 
827.44 1861.7  5171.5 15515 25858 51715 77573 N/A 
N/A 
inHg @ 0 
°
2.03602 
32.576 73.297  203.60 610.81 1018.0 2036.0 3054.0 6108.1 10180 
 
Proof pressure - maximum allowable pressure without a shift in calibration.
 
 
Burst pressure - sensor damaged or destroyed; some risk of personnel injury.