Pressure Calibrator 
 
Measure Pressure 
21 
Note 
 To ensure accuracy of the Product, the Product 
must be zeroed before a device is calibrated. 
See the “Zero Function Use” section. 
Media Compatibility 
The Product has a unique user-accessible valve cleaning 
port for easy pump service. See the “Clean the Valve 
Assembly” section for how to clean these valves. Make 
sure to only expose the Product to clean, dry gases. 
Measure Pressure with External Modules 
The Product has a digital interface to be used with 
external pressure modules. These modules are available 
in different ranges and types that include gauge, vacuum, 
differential, and absolute. Connect the modules to the 
interface and select [EXT] (external sensor). Since the 
interface between the Product and the module is digital, 
all accuracy and display resolution is derived from the 
module. See Figure 4. 
Pressure
Module
Valve
 
hix010.eps 
Figure 4. Pressure with External Modules