Fluke Calibrator, 4353234 Data Sheet

Product codes
4353234
Page of 80
 
Pressure Calibrator 
 
Measure Temperature with an RTD 
25 
Measure Temperature with an RTD 
To measure temperature with an RTD probe, select the 
RTD function on one of the displays. Make sure the 
correct probe type is selected. See the “Set the RTD 
Probe Type” section. 
The standard probe has a 10-inch insertion depth with a 
¼-inch diameter stainless steel sheath. See Figure 7. 
Note 
The factory default type is PT100-385 so if the 
Product is used with the Fluke 720 RTD Probe 
(pn 4366669), it is not necessary to set the 
probe type. Connect the probe to the Product 
and configure the display to read temperature. 
The display shows “OL” when the measured 
temperature is outside the nominal 
measurement range of the RTD function (below 
-40 
°
C or above 150 
°
C). If a custom probe is 
used, enter the R0 and coefficients with the 
serial interface (see the “Remote Operation” 
section). 
RTD Probe
 
hix016.eps 
Figure 7. Temperature Measurement with RTD Probe