Calibrator Commands
CPRT_COEFA
Use this command to entere a custom RTD into the
Product. The numeric value entered after the command is
set as the first coefficient of the polynomial used by the
custom RTD.
Product. The numeric value entered after the command is
set as the first coefficient of the polynomial used by the
custom RTD.
For example:
CPRT_COEFA 3.908300E-03
enters 3.908300e-3 as
coefficient A.
CPRT_COEFA?
Returns the number that was entered for the first
coefficient of the polynomial used in the custom RTD.
With the example above,
coefficient of the polynomial used in the custom RTD.
With the example above,
CPRT_COEFA?
returns:
3.908300E-03
CPRT_COEFB
This command is used to enter a custom RTD into the
Product. The numeric value entered after the command
will be set as the second coefficient of the polynomial
used by the custom RTD.
Product. The numeric value entered after the command
will be set as the second coefficient of the polynomial
used by the custom RTD.
For example:
CPRT_COEFB –5.774999E-07
enters –5.774999E-07
as coefficient B.
CPRT_COEFB?
Returns the number which was entered for the second
coefficient of the polynomial used in the custom RTD.
With the example above,
coefficient of the polynomial used in the custom RTD.
With the example above,
CPRT_COEFB?
returns:
-5.774999E-07
CPRT_COEFC
This command is used to enter a custom RTD into the
Product. The numeric value entered after the command
will be set as the first coefficient of the polynomial used
by the custom RTD.
Product. The numeric value entered after the command
will be set as the first coefficient of the polynomial used
by the custom RTD.
For example:
CPRT_COEFC –4.183000E-12
enters –4.183000E-12
as coefficient C.
CPRT_COEFC?
Returns the number that was entered for the third
coefficient of the polynomial used in the custom RTD. The
example above
coefficient of the polynomial used in the custom RTD. The
example above
CPRT_COEFC?
returns:
–4.183000E-12