Pressure Calibrator 
 
Menus 
21 
RESOLUTION 
To choose between a low or high resolution display: 
1. Push 
 to access the menus. 
2. Push 
 to step through the menu until 
RESOLUTION is shown on the menu bar.  
3. Push 
 to select the Resolution menu.  
4. Push 
 or  to turn low resolution on or off. 
5. Push 
 when finished. 
 
hmq062.eps 
HART 
An internal 250 
Ω
 HART resistor can be enabled when 
the Product operates in the mA Measure-24 V mode. This 
lets a HART Communicator be connected across the mA 
terminals. It is not necessary to add an external resistor. 
Note 
When the HART resistor is on, the maximum 
load-driving capacity is 750 
Ω
1. Push 
 to access the menus. 
2. Push 
 to step through the menu until HART is 
shown on the menu bar.  
3. Push 
 to select the HART menu.  
4. Push 
 or  to turn the HART Resistor on or off. 
5. Push 
 when finished. 
 
hmq063.eps