Fluke Fluke-709 Calibrator, 4234350 Data Sheet

Product codes
4234350
Page of 62
709/709H 
Users Manual 
28
 
Mode 
The Mode function from the HART. Comm. Menu lets 
you select what mode to work in. See Figure 24. 
 
gzx33.eps 
Figure 24. Mode Selection Screen 
Use the selection knob to choose the necessary mode: 
• 
mA Measure- The Product is in circuit and loop 
power is supplied externally. 
• 
mA Measure with 24V- The Product is in circuit and 
the loop power is supplied by the Product. 
• 
Communicator Only- The Product is across the 
circuit and the loop power is supplied externally. The 
250Ω resistor selection defaults to n/a (not 
applicable). 
250
Ω
 Resistor 
The 250
Ω
 Resistor function lets you turn the 250 
Ω
 
resistor on or off. Use the selection knob to change the 
resistor to the highlighted selection and return to the 
Setup and Selection screen. See Figure 25. 
 
gzx34.eps 
Figure 25. 250
Ω
 Resistor Screen