Users Manual 
26
 
Across Circuit, Communicator Only 
In Communicator Only mode, the Product is across the 
circuit and loop power is supplied externally. In Figure 22, 
when in Communicator Only mode, there must be 250 
Ω
resistance present in the loop. 
4 mA to 20 mA
RL ≥ 250 Ω
24 V
Supply
 
gzx010.eps 
Figure 22. Across Circuit, Communicator Only Connections