Precision Loop Calibrator 
 
HART Device Communication 
  
29
 
HART Connect 
The HART Connect function locates the HART device in 
the loop. Before an operation with a HART device, the 
device must be located on the loop. This is done by 
polling all of the possible device addresses and selecting 
a device from those addresses that respond to the 
search. 
If a HART protocol revision 5 or earlier device is found on 
the loop, polling stops at poll address 15. If not found, the 
Product continues to poll address 63. Polling stops after 
10 devices are found on the loop. 
If multiple devices are found on the loop, a tag list is 
shown. From the list, select the correct device. If only one 
device is found on the loop, it becomes the selected 
device by default. 
When a selected device is found, all relevant data is read 
from the device and the operations described in the 
“Function Select Menu” and “Device Setup and Data” 
sections of this manual become available. 
Polling Loop 
The Polling Loop function searches the loop for HART 
devices. This function starts immediately. The screen 
changes to show the string of dots that is extended once 
per second to as the operation progresses. See 
Figure 26. 
 
gzx35.eps 
Figure 26. Polling Screen 
The number of devices found on the loop during polling is 
shown. 
The selection knob can be pushed to stop polling early if 
it is known that all of the devices on the loop have been 
found. Push  to stop polling, disconnect from HART 
mode, and go to the function home screen. 
An error is shown if no device is found. 
If multiple devices are found, a list of tags is shown. Use 
the list of tags to select the necessary device. 
If only one device is found, the tag selection step is 
skipped.