For all ME EQuIPMENT and ME sYsTEMs
Guidance and manufacturer’s declaration – electromagnetic immunity
The SBM 50 is intended for use in the electromagnetic environment specified below. The customer or the user of the SBM 50 should
assure that it is used in such an environment.
IMMuNITY test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
material, the relative humidity should be at 
least  30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply lines Mains power quality should be that of a typi-
cal commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations 
on power supply input lines
IEC 61000-4-11
<5 %  U
T
 (>95 % dip in U
T
for 0.5 cycle
40 %  U
T
 (60 % dip in U
T
for 5 cycles
70 %  U
T
 (30 % dip in U
T
for 25 cycles
<5 %  U
T
 (>95 % dip in U
T
for 5 s
<5 %  U
T
 (>95 % dip in U
T
for 0.5 cycle
40 %  U
T
 (60 % dip in U
T
for 5 cycles
70 %  U
T
 (30 % dip in U
T
for 25 cycles
<5 %  U
T
 (>95 % dip in U
T
for 5 s
Mains power quality should be that of a typi-
cal commercial or hospital environment. If 
the user of the SBM 50 requires continued 
operation during power mains interruptions, 
it is recommended that the SBM 50 be pow-
ered from an uninterruptible power supply 
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location 
in a typical commercial or hospital environ-
ment.
NOTE:  U
T
 is the a.c. mains voltage prior to application of the test level.