Bicker Elektronik BICKER BEG-300-1224 300W PC-NETZTEIL BEG-300-1224 Data Sheet

Product codes
BEG-300-1224
Page of 2
Operating instructions
Industrial power supply 
Model: BEG-300-1224
Intended use
This power supply is used for providing power to low-voltage consumers. It is a built-in unit and therefore has to be installed in a metal housing. Only 
a qualified technician may install this unit.
Safety instructions
 
Attention, danger of life! It is strictly forbidden to modify, open, disassemble or reconvert the power supply! The BEG-300-1224 is an
 
enclosed  power  supply.  Contact  with  the  components  may  cause  an  electric  shock!   The  following  safety  label  is  to  be  affixed  to  the   
 
cabinet: „Before opening the cabinet, pull the mains plug!“ Any use other than described in these operating instructions may damage the
 
power supply and cause dangers such as short-circuit, fire etc.!
 
In case of damages due to disregard of these operating instructions any warranty will expire. Bicker will not be responsible for any consequential
 
damages! We refuse any responsibility for damages to property or persons due to inappropriate handling or disregard of these operating
 
instructions. Any warranty will expire!
Technical data
Before using this power supply, be sure to read the corresponding data sheet. If it is unavailable, you can read or download it in the internet under 
www.bicker.de. Here you will find important information such as input voltage, output rating, operating temperature, derating, connection of earth 
conductor and more. The power supply may not be operated outside of these technical data!
Installation and start-up
Only  a  qualified  technician  may  install  this  unit.  During  installation  the  respective  standards  regarding  electrical  safety  as  well  as  the  standard 
EN 60950 have to be observed. Before installation, power-off the unit by switching it off and pulling the main plug.
The PSU is a built-in unit and may preferably be used inside a metal housing. For the installation of the power supply suitable screws have to be used.
The power supply is a product according to protection class I. During installation a secure connection of the grounding equipment conductor 
between power supply and metal housing has to be ensured. The leakage current should not exceed 3,5 mA.
Cables should be run neatly and straight. Make sure the cables are not squeezed or damaged. Protect cables against sharp edges in the housing. If 
necessary, fix them using cable straps and apply a protective padding to sharp edges. Make sure the cables do not get into the computer fan.
During installation ensure that the power supply is adequately ventilated. Free circulation of air has to be possible. Also make sure that the ventilation 
slots are not covered by objects, assembly parts or ribbon cables.   
Operation
The  power  supply  may  only  be  operated  in  dry  interiors.  Contact  with  water,  even  by  condensing  air  humidity,  has  to  be  avoided  under  any 
circumstances. Operation under conditions such as extreme cold or heat, in a dusty environment and near inflammable gases, steams or solvents is 
strictly forbidden as well. Always ensure an adequate ventilation.
Derating
When operating the power supply in a higher temperature range such as +50…+70 °C the power drain has to be reduced in order to decrease the 
load applied to the unit. This is not done automatically. The user himself has to take this into consideration when setting up the system and allow for 
an adequate power reserve!
Example:  If  a  300  Watt  power  supply  has  to  be  operated  at  +60  °C  operating  temperature  and  a  derating  of  2.5  %  /  °C  within  the  range  of  
+50…+70 °C is defined in the data sheet, the output load may not exceed 225 Watts. 
  
Maintenance
The power supply does not contain any parts which need maintenance.
In case of failure, switch-off the power supply and pull the main plug. 
Repairing und therefore opening of the power supply has to be done 
by a qualified technician or repair shop.
These operating instructions are a publication of Bicker Elektronik GmbH, Ludwig-Auer-Straße 23, D-86609 Donauwörth.
Errors, mistakes as well as modifications of engineering and equipment are reserved. © 14.06.2012 by Bicker Elektronik GmbH
Disposal
 
Electrical and electronic devices may not be 
 
disposed off with domestic waste! At the end
 
of its useful life, dispose of the product 
 
according to the respective legal regulations.
Bicker Elektronik GmbH  
|  Ludwig-Auer-Straße 23  |  86609 Donauwörth  |  Germany
Tel.   +49 906 70595-0  |  Fax   +49 906 70595-55  |  info@bicker.de  |
   www.bicker.de