ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacturer’s declaration – electromagnetic immunity
The “Mini Daylight” is intended for use in the electromagnetic environment specified below. The customer
or the user of the “Mini Daylight” should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output 
lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of 
a typical commercial or hospital envi-
ronment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential
mode
±2 kV common 
mode
Mains power quality should be that of 
a typical commercial or hospital envi-
ronment.
Voltage dips, 
short interruptions 
and voltage vari-
ations on power 
supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that of a 
typical commercial or hospital environ-
ment. If the user of the “Mini Daylight” 
requires continued operation during 
power mains interruptions, it is recom-
mended that the “Mini Daylight” be 
powered from an uninterruptible power 
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical commercial 
or hospital environment.
NOTE U
T
 is the a.c. mains voltage prior to application of the test level.