Beurer ME80 User Manual

Page of 52
47
Problem
Possible cause
solution
  On launching the program, 
the message “Verification 
error” is displayed.
The device is not con-
nected to a PC or the 
Beurer ECG Manager did 
not install correctly.
Make sure that the ME 80 is 
connected to your computer 
and restart the program. If 
the verification still fails, read 
the information on installation 
on page 12 and reinstall the 
program.
9. technical Data
Model
ME 80
System of measuremen
single-channel ECG in freely selectable frontal posi-
tions/ECG signal referenced to ground (earth)
Bandwidth
0.05 to 40 Hz
Sample rate
256 Hz
Measurement range for heart 
rate:
5 to 199 beats per min.
Memory
99 measurements
Power supply
Integrated, rechargeable lithium battery
Battery capacity
Over 1,000 measurements on a full charge
Transport conditions
-20°C to +60°C, 10%–95% humidity
Storage conditions
-20°C to +60°C, 10%–95% humidity
Operating conditions
+5°C to +40°C, 30%–85% humidity
Weight
40 g