Chapter 8. Servicing the PT878
200
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
8.12.2b Installing New Software (cont.)
2.
The meter asks for confirmation. Repeat the options shown in Step 1
above. The screen on both the PC and the PT878 now appears similar to
Figure 134 below.
above. The screen on both the PC and the PT878 now appears similar to
Figure 134 below.
Figure 134: Flash Reloading Window
3.
The meter now asks if you wish to reload the flash memory via the
infrared communications port.
infrared communications port.
•
Press
[F3]
(Yes) (or C on the PC keyboard) to begin reloading via
IrCOMM, or
•
Press
[F2]
(No) (or ESC on the keyboard) to stop the reloading
procedure.
4.
Be sure the infrared sensor on the meter has a clear path to the infrared
adapter on the PC, and return to the PC Hyperterminal program.
adapter on the PC, and return to the PC Hyperterminal program.
GE Sensing PCI Loader
[HW Rev3+]
Reload Flash via Infrared (Yes/No)?
Backup Battery: OK
Firmware CRC: Valid
IrOBEX No IrCOMM
v3.0 2/26/03