Chapter 10. Specifications
214
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
Note:
Accuracy depends on pipe size and whether measurement is one-path 
or two-path. Accuracy to ±0.5% of reading may be achievable with 
process calibration.
Range:
–12.2 to 12.2 m/s (–40 to 40 ft/s)
Rangeability:
400:1
Repeatability:
 ±0.1% to 0.3% of reading
Note:
Specifications assume a fully developed flow profile (typically 10 
diameters upstream and 5 diameters downstream of straight pipe 
run) and flow velocity greater than 1 ft/s (0.3 m/s). 
Energy Accuracy
The accuracy of the energy measurement is a
(% of reading):
combination of the accuracy of the flow and
temperature measurement. Accuracy of 1% is
typical for calibrated systems.
Low Voltage
This unit complies with European Standard
Directive
EN61010-1 Installation Category II,
Compliance:
Pollution Degree 2. 
EMC
This unit complies with EN61326:1998
Directive 
CL.A Annex C. For continuous unmonitored
Compliance
operation, follow provisions of the 89/336/EMC 
Compliance.
10.2 Electrical
Internal Batteries
Type:
Rechargeable batteries. (Optional NiCad 6
R
nonrechargeable alkaline available.)
Battery Life:
9-11 hours of continuous operation is typical 
Battery Charger:
Input:
100-250 VAC, 50/60 Hz, 0.38 A
Memory:                        1 MB datalog/site battery-backed RAM
                                         1 MB program FLASH memory