Damping
The process variable (PV) Damp command changes the response time of the 
transmitter to smooth variations in output readings caused by rapid changes 
in input. Determine the appropriate damping setting based on the necessary 
response time, signal stability, and other requirements of the of loop dynamics 
of your system. The default damping value is 0.4 seconds,
(1)
 and can be reset 
to any of eleven pre-configured, nominal damping values between 0 and 25.6 
seconds.
LCD Meter Options
The Meter Options command allows you to customize the LCD meter for use 
in your application. You can configure the meter to display the 
following information:
Engineering Units
Percent of Range
User-Configurable LCD Scale
Alternating between any two of the above
The user-configurable scale is a feature that enables you to configure the 
LCD meter to a custom scale using a HART Communicator. With this feature 
you can define the decimal point position, the upper range value, the lower 
range value, the engineering units, and the transfer function. Refer to 
“Custom Meter Configuration” in Section 6 for complete configuration 
information.
DETAILED SETUP
Local Span and Zero 
Control (Local Keys)
The Local keys command allows software control over the use of the local 
span and zero adjustments. To enable or disable the span and zero 
adjustment buttons on your transmitter, perform the fast key sequence at left.
NOTE
Disabling the local keys does not disable all transmitter configuration 
changes. With the local keys disabled, you can still make changes to the 
transmitter configuration—including range values—using a HART 
Communicator.
Sensor Temperature 
Output Unit Selection
The Sensor Temperature Output Unit Selection command selects between 
Celsius and Fahrenheit units for output of the sensor temperature. The sensor 
temperature output is accessible via HART only.
DIAGNOSTICS AND 
SERVICE
The diagnostics and service functions listed here are primarily for use after 
you install the transmitter in the field. The transmitter test feature is designed 
to verify that the transmitter is operating properly, and can be performed either 
on the bench or in the field. The loop test feature is designed to verify proper 
loop wiring and transmitter output, and should only be performed after you 
install the transmitter.
HART Comm.
1, 3, 6
(1)
Rosemount 3051ND0 default damping is 3.2 seconds. Rosemount 3051ND1, with calibra-
tions below 2.5 inH
2
O (0,62 kPa), have damping set at 3.2 seconds.
HART Comm.
1, 4, 3, 4
HART Comm.
1, 4, 4, 1, 7
HART Comm.
1, 4, 1, 2, 2